FREE SHIPPING
Free shipping on all US order or order above $200
| Product Name | large-scale spray dryer |
|---|---|
| Weight | 1.000000 |
| Part No. | LBX-YMT-0024 |
| Supplier Item Code | DL411C |
| Technical Data | |
|---|---|
| Specification | Detail |
| Model | DL411C |
| Compatible Samples | Water-soluble & organic solvent (organic solvent use requires connection to GAS510C) |
| Water Evaporation Capacity | Max. 3000 mL/h |
| Spray Method | Dual-fluid nozzle (screen mesh Ø0.7mm); three-fluid nozzle method available (optional) |
| Spray/Hot Air Contact Method | Vertical downward spray, co-current flow |
| Feed Pump | 0–70 mL/min, adjustable |
| Temperature Controller Setting Range | 0–300°C |
| Temperature Control Accuracy | Inlet temp ±1°C |
| Drying Air Volume Range | 0.12–1.0 m³/min |
| Spray Air Flow Range | 0–30 L/min |
| Spray Pressure Range | 0.3–0.6 MPa |
| Drying Chamber Size | Inner diameter 457 × height 975mm (glass section) |
| Drying Chamber Material | Drying chamber, cyclone section, and product collection vessel: ultra-hard glass; other piping: stainless steel and silicone |
| External Output | Inlet/outlet temperature output (4–20mA) |
| Temperature Controller | Multi-PID control |
| Touchscreen | 7-inch, Chinese/Japanese/English; controls temperature, blower, heater, feed pump, pulse-jet switch, auto needle-clearing, alarm display, run curve |
| Control Switch | Selectable inlet or outlet temperature control mode |
| Temperature Sensor | PT100 resistance thermometer |
| Heater | 2.0 kW × 2 |
| Spray Air Pump | External spray air compressor (sold separately) or built-in compressor of GAS510C when connected (sold separately) |
| Service Outlet | For stirrer: 200–230V~, 1A |
| Intake Blower | Brushless blower |
| Filters | Intake filter, exhaust filter |
| Solvent Recovery | Requires GAS510C solvent recovery unit (sold separately) |
| Spray Nozzle Cooling | Connects to CF312L-B: 2× fittings, Ø10.5mm (sold separately) |
| Spray Air Connection | Ø7mm outer diameter fitting |
| Exhaust Connection Diameter | Ø50mm |
| Chamber Cover | Can be raised, lowered, and rotated — simplifies chamber cleaning |
| Nozzle Cooling | New automatic nozzle with cooling mechanism, standard |
| Nozzle Compatibility | Two-fluid and three-fluid nozzles supported |
| Data Logging | Experiment data recording/storage (optional) |
| Remote Control | Supported |
| Temperature Control Method | Zone-based control for faster, more stable ramp |
| Safety Functions | Inlet/outlet over-temperature protection, feed pump reverse-rotation function, overcurrent/leakage protection switch, independent over-temperature protector on heater exterior |
| External Dimensions (W×D×H) | 1080 × 850 × 1770 mm |
| Weight | ~180 kg |
| Power Supply | 200–230V~, 50/60Hz, 23–27A |
Free shipping on all US order or order above $200
Contact us 24 hours a day, 7 days a week
Simply return it within 30 days for an exchange.
We ensure secure payment with PEV